Leak Test (cont)<br />

5) To perform the Leak Test, the ventilator does the following:<br />

a) Closes the exhalation valve and sets the flow valve to a near-closed state. The display briefly<br />

shows HOMING VALVE.<br />

b) Elevates the turbine motor speed. The display shows SET TURBINE. If the display shows<br />

LEAK xx.x FAIL, see Chapter 15 - Troubleshooting for more information.<br />

c) Elevates the circuit pressure. The display shows PRES xx.x cmH 2 O where xx.x is the realtime<br />

airway pressure.<br />

d) Sets the flow valve to a near closed position. The display shows FLOW xx.x Lpm where xx.x<br />

is the flow through the flow valve.<br />

e) After several seconds, the display shows LEAK xx.x PASS or LEAK xx.x FAIL indicating the<br />

Leak Test results. The Leak Test will fail if the flow through the flow valve is greater than 1<br />

Lpm.<br />

6) To exit the Leak Test, push the Select button again and the next menu item is displayed.<br />

If the ventilator fails the Leak Test, see Chapter 15 - Troubleshooting for more information.<br />