06.02.2014 Views

LTV 1100 Operator Manual - CareFusion

LTV 1100 Operator Manual - CareFusion

LTV 1100 Operator Manual - CareFusion

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

Patient Breathing Circuit – Connection Instructions<br />

1) Connect the main breathing tube to the 22 mm outlet port on the right side of the ventilator.<br />

2) Connect the exhalation flow transducer sense line(s) to the ports marked Flow Xducer on the<br />

right side of the ventilator. These are non-interchangeable Luer fittings.<br />

3) Connect the exhalation valve drive line to the port marked Exh Valve on the right side of the<br />

ventilator.<br />

CAUTION<br />

Patient Wye Installation – After cleaning, install the patient wye in the patient circuit so the proximal<br />

sense lines are oriented up while operating.<br />

ATTENTION<br />

Installation de la soupape d'expiration - Après le nettoyage, installez la soupape d'expiration dans<br />

le circuit du patient de sorte que les lignes de détection soient alignées vers le haut pendant<br />

l’opération.<br />

C-10 <strong>Operator</strong>’s <strong>Manual</strong> - <strong>LTV</strong> ® <strong>1100</strong> Ventilator 21106-001 Rev. E

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!