LTV 1100 Operator Manual - CareFusion
LTV 1100 Operator Manual - CareFusion
LTV 1100 Operator Manual - CareFusion
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
Symptoms Possible Causes What to Do<br />
continued…<br />
Ventilator won’t<br />
exhale, repeated<br />
HIGH PRES alarms,<br />
turbine stops and<br />
pressure drops, then<br />
autocycles up to<br />
HIGH PRES again.<br />
Repeated<br />
DISC/SENSE alarms.<br />
Exhalation Drive line<br />
occluded or pinched.<br />
Internal problem with the<br />
ventilator.<br />
High or low side sense lines<br />
disconnected from vent or<br />
wye.<br />
High or low side sense line<br />
or elbow at patient wye<br />
loose or leaking.<br />
High or low sense lines are<br />
occluded.<br />
High or low sense ports in<br />
the wye are occluded.<br />
Ventilator is in Flow Sensor<br />
Mode with a Non-Flow<br />
Sensing Patient Circuit<br />
connected.<br />
Patient circuit disconnected<br />
at patient, ventilator or<br />
circuit component.<br />
Check the exhalation drive line to be sure it<br />
is correctly attached and secured at both<br />
the ventilator and exhalation valve ends.<br />
Verify the line is not occluded or pinched.<br />
Immediately contact a certified <strong>CareFusion</strong><br />
service technician.<br />
Check high and low pressure sense lines<br />
to be sure they are correctly attached and<br />
securely seated at both the ventilator and<br />
wye ends.<br />
Check the luer fitting connections for leaks.<br />
Check the elbow connectors at the wye to<br />
be sure they have not loosened or been<br />
broken loose.<br />
Verify lines are not occluded or pinched.<br />
Check the exhalation drive line at both the<br />
ventilator and exhalation valve ends. Verify<br />
the line is securely seated and not leaking.<br />
Verify that the patient circuit in use<br />
matches the ventilator patient circuit mode<br />
and verify the circuit is properly attached<br />
and passes the leak test (see Chapter 12 –<br />
Operating Procedure for additional<br />
information).<br />
Check the circuit and exhalation valve to<br />
verify the circuit is securely connected and<br />
the valve is intact. Open the exhalation<br />
valve and remove the diaphragm. Reseat<br />
the diaphragm and snap the valve cap<br />
back in place. See page 13-8 for a diagram<br />
of correct exhalation valve assembly.<br />
21106-001 Rev. E <strong>Operator</strong>’s <strong>Manual</strong> - <strong>LTV</strong> ® <strong>1100</strong> Ventilator 15-19