Control Range Tolerance<br />

Tidal Volume 58 50 to 2000 ml 10% or 10 ml, whichever is<br />

greater for temperatures from 20C<br />

to 30C only, standard atmospheric<br />

pressure<br />

Variable Flow<br />

Termination<br />

10% to 40% 15% or 2 lpm whichever is greater<br />

Variable Rise Time 1 to 9 0.1 to 1.0 sec<br />

Variable Time<br />

Termination<br />

0.3 to 3.0 sec 0.1 sec<br />

Bias Flow 0 lpm 59 , 2 lpm 60 or 10 lpm 61<br />

during exhalation<br />

SBT Start On - Off n/a<br />

10% or 1 lpm, whichever is<br />

greater<br />

SBT PS 0-30 cmH 2 O 2 cmH 2 O or 8% whichever is<br />

greater, steady-state only.<br />

SBT PEEP 0-20 cmH 2 O 2 cmH 2 O or 10%, whichever is<br />

greater<br />

SBT minutes 15-120 minutes n/a<br />

SBT high f/Vt 70-120 20%<br />

SBT low f/Vt 5-90 20%<br />

SBT High f 15-80 BPM 1 bpm or within 5%, whichever is<br />

greater<br />

SBT Low f 0-40 BPM 1 bpm or within 5%, whichever is<br />

greater<br />

Display f/Vt On - Off n/a<br />

58 For ventilator operation above 6,500 feet sea level or barometric pressures less than 605 millimeters of<br />

Mercury absolute (mmHg), see Tidal Volume in Chapter 6 – Controls for altitude and barometric pressure<br />

compensation information.<br />

59 0 lpm when O 2 Conserve ON is selected.<br />

60 2 lpm in Non-Flow Sensing Patient Circuit mode and O 2 Conserve OFF is selected.<br />

61 10 lpm in Flow Sensing Patient Circuit mode and O 2 Conserve OFF is selected.<br />