SIMV Mode<br />

SIMV mode is selected when SIMV/CPAP is selected and the Breath Rate is set between 1 and 80.<br />

In SIMV mode, machine, assist and patient breaths may be given.<br />

For the first patient trigger detected within a breath period, an assist breath is given. For all<br />

subsequent patient triggers within the same breath period, spontaneous patient breaths or pressure<br />

support if set, are given.<br />

At the beginning of a breath period, if no triggered breaths were given in the previous breath period, a<br />

machine breath is given. If there was a patient trigger in the previous breath cycle, the ventilator will<br />

not give a machine breath in the current breath period unless the set Apnea Interval is exceeded.<br />

NOTE<br />

<strong>LTV</strong> ® ventilators provide an Apnea Backup mode of ventilation. When the set Apnea Interval<br />

(maximum time allowed between the beginning of one breath and the beginning of the next breath) is<br />

exceeded, the APNEA alarm is generated and the ventilator will enter Apnea Backup ventilation<br />

mode.<br />

REMARQUE<br />

Les ventilateurs <strong>LTV</strong> ® procurent un mode de ventilation de secours pour l’apnée. Lorsque l’intervalle<br />

de l’apnée établi (durée maximum allouée entre le début d’une respiration et le début de la respiration<br />

suivante) est excédé, l’alarme APNEA est générée et le ventilateur entre en mode de ventilation de<br />

secours pour l’apnée.<br />

4-2 <strong>Operator</strong>’s <strong>Manual</strong> - <strong>LTV</strong> ® <strong>1100</strong> Ventilator 21106-001 Rev. E