06.02.2014 Views

LTV 1100 Operator Manual - CareFusion

LTV 1100 Operator Manual - CareFusion

LTV 1100 Operator Manual - CareFusion

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

Procedure for SIMV Mode Set Up<br />

Set any desired Extended Features options and:<br />

1) Push the mode Select button twice to toggle the modes between Assist/ Ctrl and SIMV/CPAP.<br />

Select the SIMV/CPAP mode.<br />

2) Establish the Breath Rate.<br />

3) Establish the Tidal Volume. The calculated peak flow Vcalc is displayed in the window while<br />

Tidal Volume is being changed.<br />

4) Establish the Inspiratory Time. The calculated peak flow Vcalc is displayed in the window while<br />

Inspiratory Time is being changed. Vcalc only applies to volume ventilation.<br />

5) Set the Pressure Support, if desired.<br />

6) Set the Sensitivity to a setting from 1 to 9.<br />

7) Set the High Pres. Limit alarm.<br />

8) Set the Low Pressure alarm.<br />

9) Set the Low Min. Vol. alarm 46 .<br />

10) Set the PEEP.<br />

46 Only available when Flow Sensing Patient Circuit mode is enabled (see Patient Circuit Mode in this Chapter<br />

for additional information).<br />

12-8 <strong>Operator</strong>’s <strong>Manual</strong> - <strong>LTV</strong> ® <strong>1100</strong> Ventilator 21106-001 Rev. E

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!