DISC/SENSE<br />

When the ventilator detects one of the following conditions, the DISC/SENSE alarm is generated:<br />

When a sense line is pinched or blocked.<br />

When a sense line has become disconnected.<br />

When a sense line is occluded (e.g., excessive condensation in the line).<br />

The ventilator detects circuit pressure during the beginning of each inspiration. If an appropriate<br />

pressure change is not detected, a DISC/SENSE alarm occurs. While the DISC/SENSE alarm is<br />

active, the ventilator cannot sense the circuit pressure so the breath is terminated.<br />

When a DISC/SENSE alarm occurs:<br />

<br />

<br />

<br />

Inspiration is immediately terminated and exhalation begins.<br />

The DISC/SENSE message is flashed in the display window.<br />

The audible alarm is sounded.<br />

To reset the DISC/SENSE alarm:<br />

1) Push the Silence Reset button to silence the alarm.<br />

2) Push the Silence Reset button to reset the alarm.<br />