Display<br />

FDw xx.xx c mH 2 0<br />

FDn xx.xx c mH 2 0<br />

FTb xx.xx Lpm<br />

LEAK xx.xx Lpm<br />

FTw xx.xx Lpm<br />

or<br />

FTn xx.xx Lpm<br />

FVd xx.xx c mH 2 0<br />

FV xx.xx Lpm<br />

STEP xxxx<br />

TS xxxx rpm<br />

BV xx.xx VOLTS<br />

EV xx.xx VOLTS<br />

PPP xx.xx c mH 2 O<br />

Real Time Data<br />

Flow differential pressure as measured at the patient wye using the<br />

wide scale transducer. Differential pressure is measured between the<br />

high and low side proximal sense lines.<br />

Flow differential pressure as measured at the patient wye using the<br />

narrow scale transducer. Differential pressure is measured between the<br />

high and low side proximal sense lines. The narrow scale transducer is<br />

only used for differential pressures between -0.35 cmH 2 O and 0.35<br />

cmH 2 O (approximately -15 Lpm to 15 Lpm).<br />

Flow in Lpm calculated from the differential pressure measured at the<br />

patient wye using the bi-directional transducer.<br />

Transducer count display is not available for this item.<br />

Leak flow calculated from the differential pressure transducer,<br />

measured at the patient wye during exhalation. This value will be<br />

approximately 0.0 when the ventilator is autocycling. Eliminate<br />

autocycling by turning the sensitivity off before reviewing this<br />

measurement.<br />

Flow in Lpm calculated from the differential pressure measured at the<br />

patient wye. When the value is calculated using the wide scale<br />

differential pressure, FTw is displayed. When the value is calculated<br />

using the narrow scale differential pressure, FTn is displayed.<br />

When Leak Compensation is on, FTw xx.xx and FTn xx.xx Lpm values<br />

are offset by the value of LEAK xx.xx Lpm.<br />

Transducer count display is not available for this item.<br />

Differential pressure as measured across the flow valve.<br />

Flow valve flow in Lpm calculated from the differential pressure<br />

measured across the flow valve.<br />

Commanded flow valve motor step position.<br />

The monitored speed of the turbine in rpms.<br />

Internal battery voltage.<br />

External power voltage.<br />

Pressure in the PEEP accumulator (PEEP Pilot Pressure)<br />