Spontaneous Breaths<br />

For Spontaneous breaths, flow is delivered to meet patient demand and maintain the circuit pressure<br />

at the measured PEEP from the previous breath. The breath is cycled when the flow drops below the<br />

Pressure Support Flow Termination setting, or below 3 lpm. Spontaneous breaths may also be<br />

terminated by exceeding 2 breath periods. Spontaneous breaths are patient type breaths.<br />

3-4 <strong>Operator</strong>’s <strong>Manual</strong> - <strong>LTV</strong> ® <strong>1100</strong> Ventilator 21106-001 Rev. E