06.02.2014 Views

LTV 1100 Operator Manual - CareFusion

LTV 1100 Operator Manual - CareFusion

LTV 1100 Operator Manual - CareFusion

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

Storage and Operating Conditions (cont.)<br />

Shock and Vibration<br />

The ventilator is designed to withstand shock and vibration in accordance with relevant<br />

requirements set forth in the following standards:<br />

IEC 68-2-27<br />

IEC 68-2-6<br />

IEC 68-2-34<br />

MIL-STD-810E<br />

Spillage<br />

Shock<br />

Vibration<br />

Vibration<br />

Shock, Ground Transport and Helicopter Transport Vibration<br />

The ventilator resists fluid spillage when tested in accordance with the relevant standards<br />

specified in IEC 601-1 Clause 44.3.<br />

External Surface Temperature<br />

External surfaces < 50C, ambient temperature of 35C n/a<br />

Communications<br />

Port Connector Specification<br />

Communications RS232, DB9 connector Protocol Options: Data, Monitor, Printer,<br />

Modem<br />

Patient Assist Call /<br />

Remote Alarm<br />

RJ11-4<br />

Closed contact resistance: 1 ohm<br />

Equipment Classification<br />

Classification The ventilator is rated as Class II equipment per IEC 601-1 Clause 6.11<br />

Type The ventilator is specified as Type BF equipment per IEC 601-1 Clause 6.11<br />

21106-001, Rev. E <strong>Operator</strong>’s <strong>Manual</strong> - <strong>LTV</strong> ® <strong>1100</strong> Ventilator A-7