HW FAULT<br />

When the ventilator detects one of the following hardware faults, the HW FAULT alarm is generated:<br />

The cooling fan is not operating, or the fan filter may be blocking the fan (see page 13-2 for<br />

cleaning and installation instructions).<br />

A problem is detected with the analog to digital converters.<br />

A problem is detected with the flow valve.<br />

A problem is detected with the processor.<br />

A problem is detected with the EEPROM memory.<br />

A problem is detected writing data to the EEPROM during system shutdown.<br />

A problem is detected with the audible alarm circuitry.<br />

A problem is detected with the alarm sounder.<br />

The HW FAULT alarm may occur as a result of ESD 33 or other transient causes. If the problem is<br />

temporary, the alarm will automatically silence when the condition clears. If the problem persists, or<br />

you experience repeated HW FAULT alarms, immediately contact a certified <strong>CareFusion</strong> service<br />

technician or <strong>CareFusion</strong>.<br />

To determine the type of hardware fault detected by the ventilator, see Appendix E - Event<br />

Trace.<br />

When a HW FAULT alarm occurs:<br />

The HW FAULT message is flashed in the display window.<br />

The audible alarm is sounded.<br />

To reset the HW FAULT alarm:<br />

1) Push the Silence Reset button twice.<br />

2) If the alarm occurs again, immediately contact a certified <strong>CareFusion</strong> service technician or<br />

<strong>CareFusion</strong>.<br />

NOTE<br />

Repeated or continuous HW FAULT alarms may indicate a hardware failure that could prevent the<br />

ventilator from performing within its specifications. Remove the ventilator from service and<br />

immediately contact a certified <strong>CareFusion</strong> service technician or <strong>CareFusion</strong>.<br />

REMARQUE<br />

Des erreurs HW FAULT répétitives ou continuelles peuvent indiquer une panne matérielle qui<br />

pourrait empêcher le ventilateur de fonctionner à l'intérieur des limites spécifiées. Retirer le<br />

ventilateur du service, et contactez immédiatement un technicien de service certifié de <strong>CareFusion</strong> ou<br />

<strong>CareFusion</strong>.<br />

33 Electrostatic Discharge.<br />

21106-001 Rev. E <strong>Operator</strong>’s <strong>Manual</strong> - <strong>LTV</strong> ® <strong>1100</strong> Ventilator 9-15