HW FAULT<br />
When the ventilator detects one of the following hardware faults, the HW FAULT alarm is generated:<br />
The cooling fan is not operating, or the fan filter may be blocking the fan (see page 13-2 for<br />
cleaning and installation instructions).<br />
A problem is detected with the analog to digital converters.<br />
A problem is detected with the flow valve.<br />
A problem is detected with the processor.<br />
A problem is detected with the EEPROM memory.<br />
A problem is detected writing data to the EEPROM during system shutdown.<br />
A problem is detected with the audible alarm circuitry.<br />
A problem is detected with the alarm sounder.<br />
The HW FAULT alarm may occur as a result of ESD 33 or other transient causes. If the problem is<br />
temporary, the alarm will automatically silence when the condition clears. If the problem persists, or<br />
you experience repeated HW FAULT alarms, immediately contact a certified <strong>CareFusion</strong> service<br />
technician or <strong>CareFusion</strong>.<br />
To determine the type of hardware fault detected by the ventilator, see Appendix E - Event<br />
Trace.<br />
When a HW FAULT alarm occurs:<br />
The HW FAULT message is flashed in the display window.<br />
The audible alarm is sounded.<br />
To reset the HW FAULT alarm:<br />
1) Push the Silence Reset button twice.<br />
2) If the alarm occurs again, immediately contact a certified <strong>CareFusion</strong> service technician or<br />
<strong>CareFusion</strong>.<br />
NOTE<br />
Repeated or continuous HW FAULT alarms may indicate a hardware failure that could prevent the<br />
ventilator from performing within its specifications. Remove the ventilator from service and<br />
immediately contact a certified <strong>CareFusion</strong> service technician or <strong>CareFusion</strong>.<br />
REMARQUE<br />
Des erreurs HW FAULT répétitives ou continuelles peuvent indiquer une panne matérielle qui<br />
pourrait empêcher le ventilateur de fonctionner à l'intérieur des limites spécifiées. Retirer le<br />
ventilateur du service, et contactez immédiatement un technicien de service certifié de <strong>CareFusion</strong> ou<br />
<strong>CareFusion</strong>.<br />