LTV 1100 Operator Manual - CareFusion
LTV 1100 Operator Manual - CareFusion
LTV 1100 Operator Manual - CareFusion
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
Appendix B - SET UP / MAINTENANCE<br />
Recommended Maintenance Schedule<br />
The <strong>LTV</strong> ® <strong>1100</strong> ventilator is designed to operate for extended periods of time with minimal routine<br />
maintenance. The following periodic maintenance is recommended:<br />
Hours of Service 68 Maintenance Required<br />
Prior to initial use Charge the Internal Battery by plugging the ventilator into an AC power source for<br />
24 hours.<br />
Setup the ventilator/accessories per Appendix C - Installation and Checkout.<br />
Check the ventilator for proper operation per Appendix C - Installation and<br />
Checkout<br />
While in storage, every<br />
two months<br />
Recharge the Internal Battery by plugging the ventilator into an AC power source<br />
for 24 hours 69 .<br />
Daily Check the Inlet Filter, clean if necessary.<br />
Check the Fan Filter, clean if necessary.<br />
If in use, a minimum of<br />
once a month<br />
Check the ventilator per Chapter 11 - Ventilator Checkout Tests. While the<br />
ventilator is off-patient, perform the Power (external) disconnect test*<br />
Verify Vte or VE monitor *<br />
Verify airway pressure or PIP monitor *<br />
* Use existing patient settings or example settings as shown in Appendix C -<br />
Installation and Checkout<br />
(Vte and VE are only available when Flow Sensing Patient Circuit mode is<br />
enabled, see Chapter 12 - Operating Procedure for additional information)<br />
Every 10,000 hours or Replace the Internal Battery 71 only with <strong>CareFusion</strong> battery P/N 18608-001. 72<br />
two years, whichever Calibrate the Transducers.<br />
comes first 70 Replace the Power Board.<br />
Replace the Sounder Assembly<br />
Clean or replace the Interior Air Inlet Filter.<br />
Every 30,000 hours or Replace the Turbine Manifold Assembly. 73<br />
six years, whichever Replace the Solenoid Manifold.<br />
comes first 70 Replace the Flow Valve.<br />
Replace the Rotary Switch Assembly.<br />
Replace the Fan Assembly.<br />
Replace all Silicone Tubing.<br />
Check the Thermo Pads for compression and replace if necessary.<br />
68<br />
To check the number of hours the ventilator has been in service, see Chapter 10 - Extended Features, Usage Meter.<br />
69<br />
If the battery is deeply discharged, it may take several charge and discharge cycles before it can be fully charged.<br />
70<br />
10,000 hour, two year and/or 30,000 hour, six year Extended Maintenance and ventilator repair must be performed by a<br />
<strong>CareFusion</strong> factory trained service technician.<br />
71<br />
Replacement at 10,000 hours or 2 years is based on normal use of up to 200 charge cycles. The battery may need to be<br />
replaced more frequently if it is being charged more often. The battery should also be replaced any time it fails to reach a full<br />
charge, or if the ventilator runs for less than ½ hour on a fully charged battery.<br />
72<br />
The <strong>LTV</strong> ® Internal Battery (P/N 18608-001) is contained in <strong>LTV</strong> ® Internal Battery Replacement Kit, P/N 18634-001.<br />
73 The Turbine Manifold Assembly (part no. 11490) must be replaced after 30,000 hours of operation. A Turbine<br />
Manifold Assembly six years or older with fewer than 30,000 hours may continue operation, as long as it<br />
passes the Performance Checkout Test (found in the <strong>LTV</strong> ® service manual) at six years and every two years<br />
thereafter (until reaching 30,000 hours). All other maintenance within the 30,000 hours, or six years of<br />
operation, is required at the specified service interval.<br />
21106-001, Rev. E <strong>Operator</strong>’s <strong>Manual</strong> - <strong>LTV</strong> ® <strong>1100</strong> Ventilator B-1