High Pressure Limit<br />

Use the High Pressure Limit to establish the maximum pressure permitted the patient circuit. When<br />

this limit is reached:<br />

A HIGH PRES alarm is displayed<br />

The audible alarm is sounded<br />

Inspiration is terminated and exhalation begins<br />

The turbine is stopped to allow the circuit pressure to evacuate when the high pressure condition<br />

persists for more than four times the set inspiratory time or more than 3.0 seconds, whichever is less.<br />

To set the High Pressure Limit:<br />

1) Push the High Pres. Limit button.<br />

2) Change the setting using the Set Value knob.<br />

Range: 5 - 100 cmH 2 O<br />

Inspiratory Time<br />

This control sets the length of the inspiratory period for Volume Controlled breaths.<br />

The Insp. Time setting, along with the Volume control setting are used to determine the peak flow for<br />

Volume controlled breaths. While the Inspiratory Time is being updated, the Calculated Peak Flow will<br />

be displayed in the display window.<br />

To set the Inspiratory Time:<br />

1) Push the Insp. Time button.<br />

2) Change the setting using the Set Value knob.<br />

Range: 0.3 - 9.9 sec<br />

6-4 <strong>Operator</strong>’s <strong>Manual</strong> - <strong>LTV</strong> ® <strong>1100</strong> Ventilator 21106-001 Rev. E