Code Event Name Event Associated Alarm<br />

37 SYNCER0 Stepper motor lost sync exited HW FAULT<br />

38 HOME ER1 Stepper motor home failure occurred HW FAULT<br />

39 HOME ER0 Stepper motor home failure exited HW FAULT<br />

40 EEPROM EEPROM degraded HW FAULT<br />

41 CRC Memory CRC check failed RESET<br />

42 HI PRES1 High pressure occurred HIGH PRES<br />

43 HI PRES0 High pressure exited HIGH PRES<br />

44 TBN ISTP Turbine immediate stop occurred HIGH PRES<br />

45 TBN ZERO Turbine set to zero flow occurred HIGH PRES<br />

46 TBN ESTP Turbine emergency stop occurred HIGH PRES<br />

47 LOW VE 1 Low minute volume occurred LOW MIN VOL<br />

48 LOW VE 0 Low minute volume exited LOW MIN VOL<br />

49 LO PRES1 Low peak pressure occurred LOW PRES<br />

50 LO PRES0 Low peak pressure exited LOW PRES<br />

51 CLR EVNT Event log cleared N/A<br />

52 CLR CTRL Control settings cleared N/A<br />

53 SET DATE Date set N/A<br />

54 SET TIME Time set N/A<br />

55 N/A<br />

56 STACK Stack overflow detected RESET<br />

57 POST POST failure RESET<br />

58 RUNAWAY Code runaway detected RESET<br />

59 WDOG TST Watchdog test run Inop<br />

60 CLR CAL Calibration records cleared N/A<br />

61 XDCR NAR<br />

62 XDCR WID<br />

63 XDCR BI<br />

Differential pressure transducer -<br />

Narrow channel fault<br />

Differential pressure transducer -<br />

Wide channel fault<br />

Differential pressure transducer - Bidirectional<br />

channel fault<br />

XDC FLT1<br />

XDC FLT1<br />

XDC FLT1<br />

64 XDCR AIR Airway pressure transducer fault XDC FLT1<br />

65 ADC1 VAL<br />

AD mismatch primary channel fault<br />

value<br />

HW FAULT<br />

66 TBN HSTP Turbine Hold Stop occurred HIGH PRES<br />

67 LN VENT1<br />

68 FLUSH ER<br />

69 RAC ERR1<br />

70 RAC ERR0<br />

Shutdown for other than pressing<br />

On/Standby button<br />

A problem is detected writing data to<br />

the EEPROM during system<br />

shutdown.<br />

Problem detected with primary and/or<br />

redundant audible alarm circuitry<br />

Recovery from problem detected with<br />

primary and/or redundant audible<br />

alarm circuitry<br />

RESET<br />

HW FAULT<br />

HW FAULT<br />

HW FAULT<br />

71 SNDRERR1 Alarm sounder error HW FAULT<br />

E-4 <strong>Operator</strong>’s <strong>Manual</strong> - <strong>LTV</strong> ® <strong>1100</strong> Ventilator 21106-001 Rev. E