Alarms<br />

Variable Alarms<br />

Control Range Tolerance<br />

Apnea Interval 10 to 60 seconds 0.5 seconds<br />

High Breath Rate Rate: 5 - 80 bpm - HIGH f OFF 1 bpm or within 5% of breath<br />

period, whichever is greater.<br />

Time: 0 - 60 sec<br />

0.1 seconds<br />

High PEEP 3 to 20 cmH 2 O above set PEEP 2 cmH 2 O or 10%, whichever is<br />

greater.<br />

Low PEEP<br />

-3 to - 20 cmH 2 O below set<br />

PEEP<br />

2 cmH 2 O or 10%, whichever is<br />

greater.<br />

High Pressure Limit 5 to 100 cmH 2 O 5 to 20 cmH 2 O: 2 cmH 2 O<br />

21 to 100 cmH 2 O: 4 cmH 2 O<br />

HP Alarm Delay No Delay, 1 Breath, 2 Breaths Only audible portion of alarm<br />

notification is delayed.<br />

Low Minute Volume 0.1 to 99 liters 15% or the measured total breath<br />

rate times 15 ml, whichever is<br />

greater.<br />

Low Peak Pressure “- -”, 1 to 60 cmH 2 O 2 to 20 cmH 2 O: 2 cmH 2 O<br />

21 to 60 cmH 2 O: 4 cmH 2 O<br />

LPP Alarm All Breaths, VC/PC Only Select breath types Low Pressure<br />

alarm applies to.<br />

SBT High f/Vt Off, 70 – 900 f/Vt 2 cmH 2 O or 10%, whichever is<br />

greater<br />

SBT Low f/Vt Off, 5 – 90 f/Vt 2%<br />

SBT High f Off, 15 – 80 bpm 1 bpm or within 5% of the breath<br />

period, whichever is greater<br />

SBT Low f Off, 1 – 40 bpm 1 bpm or within 5% of the breath<br />

period, whichever is greater<br />

21106-001, Rev. E <strong>Operator</strong>’s <strong>Manual</strong> - <strong>LTV</strong> ® <strong>1100</strong> Ventilator A-3