RESET / RESET 1<br />

A RESET or a RESET 1 alarm occurs if the ventilator restarts following a condition other than being<br />

shut down by pressing the On/Standby button.<br />

The ventilator runs an ongoing set of self-tests to verify that it is operating correctly. If the ventilator<br />

detects a condition that makes safe ventilator operation uncertain, it reinitializes itself to allow the<br />

more sophisticated Power On Self Tests (POST) to be performed. If the POST does not detect any<br />

further problems, the ventilator will resume operation and a RESET or a RESET 1 alarm is posted. If<br />

the POST detects a problem that could cause continued operation to be unsafe, a ventilator INOP will<br />

occur.<br />

Conditions that could cause a RESET or a RESET 1 alarm:<br />

Operating the ventilator on the internal battery until it is fully depleted.<br />

Electrostatic Discharge (ESD)<br />

Other transient causes.<br />

When a RESET or a RESET 1 alarm is generated:<br />

An error code is written to the Event Trace indicating the type of problem detected.<br />

The ventilator resets itself and performs the Power On Self Tests (POST).<br />

If no further problems are detected, the ventilator resumes operation.<br />

The RESET or RESET 1 message is flashed in the display window.<br />

The audible alarm is sounded.<br />

To reset the RESET or a RESET 1 alarm:<br />

1) Push the Silence Reset button twice.<br />

NOTE<br />

When a RESET or RESET 1 alarm has occurred, check the Event Trace for more information about<br />

the problem. See Appendix E - Event Trace for more information about events.<br />

Repeated occurrences of the RESET or RESET 1 alarm may indicate a problem with the ventilator’s<br />

hardware. Please immediately contact a certified <strong>CareFusion</strong> service technician.<br />

REMARQUE<br />

Lorsqu'une alarme RESET ou RESET 1 se produit, vérifiez le suivi de l'événement pour obtenir plus<br />

d'informations sur le problème. Consultez l'Annexe E - Suivi de l'événement, pour plus d'informations<br />

à propos de l'événement.<br />

Des occurrences répétitives de l'alarme RESET ou RESET 1 peuvent indiquer un problème matériel<br />

avec le ventilateur. Veuillez contacter immédiatement un technicien de service certifié de <strong>CareFusion</strong>.<br />

9-24 <strong>Operator</strong>’s <strong>Manual</strong> - <strong>LTV</strong> ® <strong>1100</strong> Ventilator 21106-001 Rev. E