Chapter 12 - OPERATING PROCEDURE<br />
This section describes how to turn the <strong>LTV</strong> ® <strong>1100</strong> ventilator on and off, and how to set up the<br />
ventilation modes.<br />
NOTE<br />
In the absence of an external power source, the ventilator automatically begins operation using the<br />
internal battery. Do not operate the <strong>LTV</strong> ® <strong>1100</strong> exclusively on the internal battery as a standard operating<br />
practice. The internal battery should be used for emergency situations only or for short periods while<br />
switching between external power supply connections.<br />
REMARQUE<br />
S’il n’y a pas de source d’alimentation externe, le ventilateur s’alimente automatiquement de la pile<br />
interne. Ne pas utiliser le <strong>LTV</strong> ® <strong>1100</strong> exclusivement avec la batterie interne en tant que procédure<br />
d’exploitation normale. La batterie interne doit être réservée aux situations d’urgence seulement ou<br />
pour de courtes périodes de temps pendant le transfert entre des sources d’alimentation externes.<br />
Patient Circuit Mode<br />
The <strong>LTV</strong> ® <strong>1100</strong> ventilator features two patient circuit modes to accommodate a flow sensing patient<br />
circuit or a non-flow sensing patient circuit. The patient circuit mode is selected upon startup and only<br />
when the patient circuit query is turned on (see Queries in Chapter 10 - Extended Features for<br />
additional information).<br />
<br />
<br />
Flow Sensing Patient Circuit Mode (FLOW SENSOR)<br />
The Flow Sensing Patient Circuit mode is intended to be used with a flow sensing patient circuit<br />
and provides features such as: VE and Vte monitoring, Low Minute Volume alarm, patient flow<br />
trigger sensitivity (with a bias flow of 10 lpm), and Spontaneous Breathing Trial (SBT).<br />
Non-Flow Sensing Patient Circuit Mode (NO FL SENSOR)<br />
The Non-Flow Sensing Patient Circuit mode is intended to be used with a non-flow sensing<br />
patient circuit and provides pressure trigger sensitivity (with a bias flow of 2 lpm) with basic<br />
monitoring and alarm functions. VE and Vte monitoring, and the Low Minute Volume alarm are<br />
not available in this mode.<br />
WARNING<br />
Patient Circuit Mode – The ventilator must be set to the applicable patient circuit mode, based on<br />
the patient circuit type in use. A patient circuit type and ventilator mode mismatch will result in a<br />
DISC/SENSE alarm or VE monitoring, Vte monitoring, Low Minute Volume alarm, and SBT features<br />
unavailable. To avoid unintended alarms and/or receive desired monitored values, be sure the correct<br />
patient circuit is used and the ventilator patient circuit mode is set appropriately.<br />
21106-001 Rev. E <strong>Operator</strong>’s <strong>Manual</strong> - <strong>LTV</strong> ® <strong>1100</strong> Ventilator 12-1