Symptoms Possible Causes What to Do<br />
continued…<br />
Monitored volume is<br />
low.<br />
Delivered volume is<br />
low.<br />
Delivered volume is<br />
half the set volume.<br />
Delivered pressure is<br />
low, PEEP is low,<br />
ventilator is<br />
autocycling.<br />
Delivered pressure is<br />
low.<br />
Monitored pressure is<br />
low.<br />
Exhalation drive line<br />
leaking or loose.<br />
Exhalation valve<br />
leaking during<br />
inspiration.<br />
Sense lines are<br />
reversed.<br />
Leak Compensation is<br />
not on.<br />
Failed autozero.<br />
Failed calibration or<br />
internal problem with<br />
the ventilator.<br />
VHome setting does<br />
not match flow valve.<br />
Circuit leak.<br />
Check the exhalation drive line at both the<br />
ventilator and exhalation valve ends. Verify the<br />
line is securely seated and not leaking.<br />
Verify the exhalation valve is not leaking during<br />
inspiration. If it is leaking, open the exhalation<br />
valve and remove the diaphragm. Reseat the<br />
diaphragm. See page 13-8 for a diagram of<br />
correct exhalation valve assembly.<br />
If necessary, replace the exhalation diaphragm,<br />
or exhalation valve with a new one.<br />
The sense lines are not designed to be removed<br />
from either the wye or the luer fittings. If the<br />
sense lines have been removed and replaced<br />
incorrectly, they may not seal correctly when<br />
replaced. Replace the patient wye and sense<br />
lines with a known good assembly.<br />
Verify that the Leak Compensation extended<br />
features option is set to On (default setting is on).<br />
See page 10-12 for instructions.<br />
Perform an autozero under XDCR ZERO. See<br />
page 10-27 for more information.<br />
Immediately contact a certified <strong>CareFusion</strong><br />
service technician.<br />
Immediately contact a certified <strong>CareFusion</strong><br />
service technician.<br />
Run a leak test and reseat or replace the leaking<br />
parts or connections. See page 11-8 for<br />
instructions.<br />
15-8 <strong>Operator</strong>’s <strong>Manual</strong> - <strong>LTV</strong> ® <strong>1100</strong> Ventilator 21106-001 Rev. E