Appendix A - VENTILATOR SPECIFICATIONS<br />

Modes and Breath Types<br />

Breath Types<br />

Modes<br />

Volume Control, Pressure Support, Spontaneous<br />

Control, Assist/Control, SIMV, CPAP, NPPV, Apnea Backup<br />

Variable Controls<br />

Control Range Tolerance<br />

Backup Pressure Trigger -3 cmH 2 O 2 cmH 2 O<br />

Breath Rate “--”, 1 to 80 bpm 1 bpm or 10% of breath period,<br />

whichever is less<br />

Date Format<br />

Display Select<br />

mm/dd/yyyy, dd/mm/yyyy,<br />

yyyy/mm/dd<br />

Toggles between manual or<br />

automatic display scrolling<br />

and changes monitor<br />

displayed.<br />

Inspiratory Time 0.3 to 9.9 seconds 0.05 seconds<br />

n/a<br />

n/a<br />

Leak Compensation On, Off n/a<br />

Language<br />

Patient Circuit Mode<br />

Query<br />

English, Dansk, Deutsch,<br />

Espanol, Francais,<br />

Italiano, Norsk, Portugues,<br />

Svenska, Русско<br />

On, Off<br />

n/a<br />

n/a<br />

Leak Query On, Off n/a<br />

PEEP/CPAP 0 to 20 cmH 2 O Uncalibrated<br />

PIP LED Display On, Off n/a<br />

Pressure Support “--”, 1 to 60 cmH 2 O 2 cmH 2 O or 8% whichever is<br />

greater, steady-state only.<br />

Set Date 1/1/1998 to 12/31/2097 n/a<br />

Set Time 00:00:00 to 23:59:59 n/a<br />

Sensitivity:<br />

Flow Sensing Mode:<br />

Non-Flow Sensing<br />

Mode / Oxygen<br />

Conserve On:<br />

1 to 9 Lpm, “-”<br />

1 to 9 cmH 2 O, “-”<br />

+ 1/- 0.5 lpm for setting of 1; 1<br />

lpm for all other settings.<br />

2 cmH 2 O or 10% of measured<br />

pressure, whichever is greater<br />

21106-001, Rev. E <strong>Operator</strong>’s <strong>Manual</strong> - <strong>LTV</strong> ® <strong>1100</strong> Ventilator A-1