NO CAL DATA, NO CAL Monitor Display<br />

When the ventilator detects invalid or missing calibration records on power up, the NO CAL DATA<br />

alarm is generated. When this happens, default calibration values are used, and although the<br />

ventilator will continue to operate, the accuracy of volumes and pressures may be reduced.<br />

A NO CAL message is posted in place of affected monitored values when the ventilator is operating<br />

without valid transducer calibration data.<br />

When the NO CAL DATA alarm occurs:<br />

The NO CAL DATA message is flashed in the display window.<br />

The audible alarm is sounded.<br />

The ventilator continues to operate.<br />

Default transducer data is used.<br />

Vte, PIP, MAP, PEEP, and VE monitored values are displayed as NO CAL.<br />

To clear the NO CAL DATA alarm:<br />

1) Push the Silence Reset button twice. This will clear the alarm and the ventilator will continue to<br />

operate; however, the NO CAL message will still be displayed in place of affected monitored<br />

values.<br />

2) Take the unit out of service and perform the Calibration procedure 35 .<br />

To clear the NO CAL message:<br />

1) Take the unit out of service and perform the Calibration procedure 35 .<br />

WARNING<br />

NO CAL Condition - Operation of the <strong>LTV</strong> ® <strong>1100</strong> ventilator under a NO CAL condition may result in<br />

inaccurate pressure and volume measurements. Should this condition occur, disconnect the patient<br />

from the ventilator, provide an alternative method of ventilation and immediately contact a certified<br />

<strong>CareFusion</strong> service technician or <strong>CareFusion</strong>.<br />

AVERTISSEMENT<br />

Condition NO CAL - L'opération continue du ventilateur de la <strong>1100</strong> <strong>LTV</strong> ® sous condition NO CAL<br />

peut résulter en mesures de pression et de volume erronées. Si cette condition se présente, le<br />

ventilateur doit être retiré du service, et vous devez immédiatement contacter votre technicien de<br />

service certifié de <strong>CareFusion</strong> ou <strong>CareFusion</strong>.<br />

35 For more information on the Calibration procedure, see the <strong>LTV</strong> ® 1200, 1150, and <strong>1100</strong> Ventilator Service<br />

<strong>Manual</strong>, p/n 18603-001.<br />

9-20 <strong>Operator</strong>’s <strong>Manual</strong> - <strong>LTV</strong> ® <strong>1100</strong> Ventilator 21106-001 Rev. E