Procedure for CPAP Mode Set Up<br />

Set any desired Extended Features options and:<br />

1) Push the mode Select button twice to toggle the modes between Assist/ Ctrl and SIMV/CPAP.<br />

Select the SIMV/CPAP mode.<br />

2) Set the Breath Rate to dashes “- -”.<br />

3) Establish the Tidal Volume for apnea backup. The calculated peak flow Vcalc is displayed in<br />

the window while Tidal Volume is being changed.<br />

4) Establish the Inspiratory Time for apnea backup. The calculated peak flow Vcalc is displayed in<br />

the window while Inspiratory Time is being changed. Vcalc only applies to volume ventilation.<br />

5) Set the Pressure Support, if desired.<br />

6) Set the Sensitivity to a setting from 1 to 9.<br />

7) Set the High Pres. Limit alarm.<br />

8) Set the Low Pressure alarm for apnea backup.<br />

9) Set the Low Min. Vol. alarm 47 .<br />

10) Set the PEEP.<br />

47 Only available when Flow Sensing Patient Circuit mode is enabled (see Patient Circuit Mode in this Chapter<br />

for additional information).<br />

21106-001 Rev. E <strong>Operator</strong>’s <strong>Manual</strong> - <strong>LTV</strong> ® <strong>1100</strong> Ventilator 12-9