Ventilator Proper Operation Worksheet<br />

SERIAL NUMBER: ________________ CONDUCTED BY: _______________________<br />

DATE: ________________<br />

TEST<br />

DESCRIPTION<br />

PAGE /<br />

STEP<br />

Ventilator Checkout Tests (Chapter 11 -)<br />

MEAS.<br />

VALUE<br />

REQUIREMENT PASS /<br />

FAIL<br />

Alarm Test 11-3 Audible alarm must activate for minimum 2<br />

sec’s.<br />

Confirming audible Chirp must activate<br />

after alarm is silenced<br />

Display Test 11-4 All displays must light except Vent Inop<br />

Control Test 11-6 Correct messages displayed in window<br />

Leak Test 11-8 “X.X PASS”, Record value displayed<br />

Vent Inop Alarm Test 11-10 Alarm sounded and Inop LED illuminated<br />

15 sec’s.<br />

Confirming audible Chirp must activate<br />

after alarm is silenced<br />

Checking the Ventilator for Proper Operation (Appendix C - Installation and Checkout):<br />

Ventilator Settings:<br />

Settings:<br />

Patient Circuit Mode: Flow Sensor<br />

Mode: Assist/Ctrl<br />

Breath Rate: 12<br />

Tidal Volume: 500<br />

Insp. Time: 1 sec<br />

Pressure Support: 0<br />

Sensitivity: 3<br />

High Pressure Limit: 100<br />

Low Pressure Alarm: 5<br />

Low Min Vol: 1.0<br />

PEEP: 0<br />

C-19<br />

2) A)<br />

Selected Monitors should read as follows:<br />

Exhaled Tidal Volume: 383 to 633 ml<br />

I:E Ratio : 1:3.8 to 1:4.2<br />

Total Breath Rate: 12 bpm<br />

Total Minute Vol: 4.6 to 7.6 L<br />

No alarms<br />

21106-001, Rev. E <strong>Operator</strong>’s <strong>Manual</strong> - <strong>LTV</strong> ® <strong>1100</strong> Ventilator C-21