SBT Operations (cont.)<br />

4) Turn the Set Value knob until MINUTES is displayed, push the Select button, and xxx MIN is<br />

displayed.<br />

Turn the Set Value knob until the desired setting is displayed, push the Select button, and the<br />

SBT ventilation mode run time is set.<br />

Range: 15 - 120 minutes, in increments of 5<br />

5) Turn the Set Value knob until HIGH f/Vt is displayed, push the Select button, and xxx f/Vt is<br />

displayed.<br />

Turn the Set Value knob until the desired setting is displayed, push the Select button, and the<br />

SBT high f/Vt alarm value is set (see SBT > f/Vt in Chapter 9 - Ventilator Alarms, for additional<br />

information).<br />

Range: HI f/Vt OFF or 70 - 900 f/Vt, in increments of 5<br />

6) Turn the Set Value knob until LOW f/Vt is displayed, push the Select button, and xxx f/Vt is<br />

displayed.<br />

Turn the Set Value knob until the desired setting is displayed, push the Select button, and the<br />

SBT low f/Vt alarm value is set (see SBT < f/Vt in Chapter 9 - Ventilator Alarms, for additional<br />

information).<br />

Range: LO f/Vt OFF or 5 - 90 f/Vt, in increments of 5<br />

7) Turn the Set Value knob until SBT HIGH f is displayed, push the Select button, and xxx bpm is<br />

displayed.<br />

Turn the Set Value knob until the desired setting is displayed, push the Select button, and the<br />

SBT high f alarm value is set (see SBT > f in Chapter 9 - Ventilator Alarms, for additional<br />

information).<br />

Range: SBT HI f OFF or 15 - 80 bpm, in increments of 1<br />

8) Turn the Set Value knob until SBT LOW f is displayed, push the Select button, and xxx bpm is<br />

displayed.<br />

Turn the Set Value knob until the desired setting is displayed, push the Select button, and the<br />

SBT low f alarm value is set (see SBT < f in Chapter 9 - Ventilator Alarms, additional<br />

information).<br />

Range: SBT LO f OFF or 1 - 40 bpm, in increments of 1<br />

9) Turn the Set Value knob until DISPLAY f/Vt is displayed, push the Select button, and DISPLAY<br />

OFF or DISPLAY ON is displayed.<br />

Turn the Set Value knob until the desired setting is displayed, push the Select button, and the<br />

SBT scrolling display status (on or off) of the monitored xxx f/Vt value is set (see Automatic or<br />

<strong>Manual</strong> Data Display Scrolling in Chapter 8 - Monitored Data, for additional information).<br />

Exit<br />

To return to the top of the SBT OP menu:<br />

1) Push the Select button while EXIT is displayed.<br />

21106-001 Rev. E <strong>Operator</strong>’s <strong>Manual</strong> - <strong>LTV</strong> ® <strong>1100</strong> Ventilator 10-25