Leak Test<br />

Use the Leak Test to test the patient circuit for leaks.<br />

To run the Leak Test:<br />

1) Attach all patient circuit accessories (such as water traps, heated circuits and humidifiers) to the<br />

patient circuit.<br />

2) Connect the patient circuit to the ventilator.<br />

3) With a clean, gloved hand or 4”X4” gauze pad, occlude the proximal end of the patient circuit.<br />

4) Push the Select button while LEAK is displayed.<br />

NOTE<br />

The Leak Test cannot be run until the ventilator has been running for 60 seconds. If you attempt to<br />

run the leak test before the warm-up period has completed, a WARMUP xx message will be<br />

displayed. When the warm-up period is complete, the Leak Test menu item is redisplayed.<br />

REMARQUE<br />

Le test de fuites ne peut s'exécuter tant que le ventilateur n'a pas fonctionné durant 60 secondes.<br />

Si vous tentez d'exécuter un test de fuites avant que la période de réchauffement ne soit<br />

complétée, un message «WARMUP xx» sera affiché. Lorsque la période de réchauffement est<br />

complétée, les éléments du menu Test de fuites sont de nouveau affichés.<br />

11-8 <strong>Operator</strong>’s <strong>Manual</strong> - <strong>LTV</strong> ® <strong>1100</strong> Ventilator 21106-001 Rev. E