PEEP Control<br />

The PEEP control establishes the Positive End Expiratory Pressure.<br />

To set the PEEP:<br />

1) Push the PEEP control button.<br />

2) Change the setting using the Set Value knob.<br />

Range: 0 – 20 cmH 2 O<br />

The <strong>LTV</strong> ® <strong>1100</strong> has PEEP Compensated Pressure Support.<br />

On the left waveform (breath), the ventilator is set to deliver Pressure Support of 20 cmH 2 O with no<br />

PEEP. The “span” or “delta” is 20 cmH 2 O.<br />

On the right waveform, the vent is set to deliver Pressure Support of 20 cmH 2 O. Note that 5 cmH 2 O of<br />

PEEP have been added. This ventilator “compensates” for the added PEEP by automatically<br />

increasing the Pressure Support in order to maintain the same “span” or “delta” of 20 cmH 2 O.<br />

To view the measured PEEP in the display window:<br />

1) Use the Select button to display the measured PEEP in the display window.<br />

6-10 <strong>Operator</strong>’s <strong>Manual</strong> - <strong>LTV</strong> ® <strong>1100</strong> Ventilator 21106-001 Rev. E