To clean the exhalation valve, sense lines, wye and reusable patient circuit:<br />

1) Remove all particulate matter and bathe for a minimum of 10 minutes using one of the following<br />

solutions warmed to 95°F to 150°F (35°C to 65.5°C):<br />

mild detergent, 50% water / 50% white vinegar solution, or liquid cleaner (e.g., KlenZyme®<br />

or other enzymatic cleaner)<br />

Ultrasonic cleaning is not recommended.<br />

2) Rinse gently for 2 minutes and use a low flow air source (less than 10 L/min) to eliminate any<br />

residual fluid.<br />

To clean the water trap (if used), use a mild detergent solution followed by rinsing and drying<br />

with a low-flow air source.<br />

To high level disinfect the exhalation valve, sense lines, wye and reusable patient circuit:<br />

1) Remove all particulate matter and bathe in a glutaraldehyde solution (e.g., Cidex (2%)) for 20<br />

minutes.<br />

2) Rinse gently for 2 minutes and use a low flow air source (less than 10 L/min) to eliminate any<br />

residual fluid.<br />

Sterilization:<br />

Sterilization of the patient circuit components and water traps should follow individual institution<br />

processes or guidelines.<br />

To reassemble the exhalation valve and patient circuit:<br />

1) Insert the side of the diaphragm with the scalloped edge into the mating cavity on the inside of<br />

the valve cap.<br />

CAUTION<br />

Valve assembly - If the diaphragm is inserted backwards when assembling the valve, it may<br />

adversely affect the operation of the ventilator. Be sure to insert the side of the diaphragm with the<br />

scalloped edge into the mating cavity on the inside of the valve cap.<br />

ATTENTION<br />

Assemblage de la valve – Si le diaphragme est inséré à l’envers lors de l’assemblage de la valve, il<br />

peut entraver le fonctionnement du ventilateur. S’assurer d’insérer le côté du diaphragme dont le bord<br />

est festonné dans la cavité commune à l’intérieur du couvercle de la valve.<br />

13-8 <strong>Operator</strong>’s <strong>Manual</strong> - <strong>LTV</strong> ® <strong>1100</strong> Ventilator 21106-001 Rev. E