Display Test (cont)<br />

Verify displays are illuminated in the following colors:<br />

Display Color Display Color<br />

Airway Pressure Display Green SIMV/CPAP Mode LED Green<br />

Display Window Red NPPV Mode LED Green<br />

Breath Rate Green <strong>Manual</strong> Breath LED Green<br />

Tidal Volume Green Control Lock LED Green<br />

Inspiratory Time Green PEEP Green<br />

Pressure Support Green Patient Effort LED Green<br />

Sensitivity Green External Power LED Not tested<br />

High Pressure Limit Alarm Red Charge Status LED Not tested<br />

Low Pressure Alarm Red Battery Level LED Amber<br />

Low Minute Volume Alarm Red Vent Inop LED Not tested<br />

On/Standby LED Green Silence Reset LED Red<br />

Assist/Control Mode LED Green<br />

3) To end the display test, push the Select button again and the next menu item is displayed.<br />

If the Display Test fails, see Chapter 15 - Troubleshooting for more information.<br />