HIGH PEEP<br />

When the patient circuit positive end expiratory pressure (PEEP) exceeds the High PEEP alarm<br />

setting, the HIGH PEEP alarm is generated.<br />

When a HIGH PEEP alarm occurs:<br />

The HIGH PEEP message is flashed in the display window.<br />

The audible alarm is sounded.<br />

To reset the HIGH PEEP alarm:<br />

1) Push the Silence Reset button once to silence the audible alarm.<br />

2) Push the Silence Reset button twice to reset the alarm (silences the audible alarm and clears the<br />

flashing display).<br />

9-12 <strong>Operator</strong>’s <strong>Manual</strong> - <strong>LTV</strong> ® <strong>1100</strong> Ventilator 21106-001 Rev. E