DEFAULTS (cont.)<br />
The factory-set default Control settings are:<br />
See Chapter 6 - Controls, for detailed information concerning specific Controls.<br />
Control - Default<br />
Breath Rate - 12 bpm<br />
Control Lock - On<br />
Data Display Scrolling - Auto-On<br />
High Pressure Limit - 20 cmH 2 O<br />
Inspiratory Time - 1.5 sec<br />
Low Minute Volume - 2.5 Lpm<br />
Control - Default<br />
Low Pressure - 5 cmH 2 O<br />
PEEP - 0 cmH 2 O<br />
Pressure Support - 1 cmH 2 O<br />
Sensitivity - 2 Lpm<br />
Tidal Volume - 500 ml<br />
Ventilation Mode - Assist/Control<br />
The factory-set default Extended Features settings are:<br />
See Chapter 10 - Extended Features, for detailed information concerning specific Features.<br />
Feature - Default<br />
Feature - Default<br />
Alarm Volume - 85 dBA<br />
Patient Circuit Query - Off<br />
Apnea Interval - 20 sec<br />
Patient Circuit Mode - Flow Sensor<br />
Com Setting 29 - Data<br />
PIP LED - On<br />
Control Unlock - Easy Rise Time Profile - 4<br />
Date Format - MM/DD/YYYY<br />
SBT Display f/Vt Alarm - Off<br />
High f Alarm - High f Off<br />
SBT High f Alarm - 35 bpm<br />
High f Alarm Delay - 30 sec<br />
SBT High f/Vt Alarm - 105 f/Vt<br />
High PEEP Alarm - PEEP +5 cmH 2 O<br />
SBT Low f Alarm - 10 bpm<br />
HP Alarm Delay - No Delay<br />
SBT Low f/Vt Alarm - 70 f/Vt<br />
Language - English<br />
SBT Mode - Off<br />
Leak Compensation - On<br />
SBT Mode Run Time - 20 min<br />
Low PEEP Alarm - PEEP -5 cmH 2 O<br />
SBT PEEP - 0 cmH 2 O<br />
LPP Alarm - All Breaths<br />
SBT Pressure Support - 10 cmH 2 O<br />
NPPV Mode - Off Variable Flow Termination - 25%<br />
Patient Assist - Normal<br />
Variable Time Termination - 1.5 sec<br />
O 2 Conserve - Off<br />
Sigh Breath - Off<br />
Leak Test Query - Off<br />
29 This feature is not reset to default values when the SET DEFAULTS option is used in Extended Features.<br />