Mechanical Controls<br />

Control Range Tolerance<br />

Over Pressure Relief ≤125 cmH 2 O N/A<br />

Sub-Ambient Relief Pressure Drop: 5 cmH 2 O at 50 lpm<br />

Internal Compliance<br />

Compliance<br />

< 0.1 mL/cm<br />

Monitors<br />

Monitor Range Tolerance<br />

Calculated Peak Flow 10 to 100 lpm 2 lpm or 10%, whichever is greater<br />

Exhaled Tidal Volume 0 to 4000 ml 15% or 15 ml, whichever is greater<br />

I:E Ratio, Measured 99:1 and 1:99<br />

Based on the measured<br />

inspiratory / exhalation times<br />

I:E Ratio, Calculated<br />

1:99 to 4.0:1 based on set<br />

breath rate and inspiratory<br />

time<br />

Accuracy for times are 50 ms or 5%,<br />

whichever is greater<br />

5%<br />

Mean Airway Pressure 0 to 99 cmH 2 O 2 cmH 2 O or 10%, whichever is greater<br />

Peak Inspiratory<br />

Pressure<br />

0 to 120 cmH 2 O 2 cmH 2 O or 5%, whichever is greater<br />

PEEP 0 to 99 cmH 2 O 2 cmH 2 O or 10%, whichever is greater<br />

Total Breath Rate 0 to 250 breaths per minute 1 bpm or within 5% of the breath<br />

period, whichever is greater<br />

Total Minute Volume 0 to 99.9 liters 15%, or the measured total breath rate<br />

times 15 ml, whichever is greater<br />

SBT Minutes 15 – 120 minutes 0.1 seconds<br />

f/Vt<br />

and<br />

f<br />

0 – 4000<br />

0 – 250bpm<br />

20%<br />

Total breath rate: 1 bpm or within 5%<br />

of breath period, whichever is greater<br />

Button Controls<br />

Display<br />

Control Lock<br />

<strong>Manual</strong> Breath<br />

Standby/On<br />

Silence Reset<br />

Function<br />

Locks front panel controls, can be set to Easy or Hard unlocking<br />

Generates a machine breath<br />

Puts ventilator in On or Standby state<br />

Silences and resets alarms<br />

21106-001, Rev. E <strong>Operator</strong>’s <strong>Manual</strong> - <strong>LTV</strong> ® <strong>1100</strong> Ventilator A-5