Before Connecting the Ventilator to a Patient<br />

Perform the following steps before connecting the ventilator to a patient:<br />

1) If this is the initial use of the ventilator, follow the checkout procedures in Appendix C -<br />

Installation and Checkout before proceeding.<br />

2) If desired, you may connect the ventilator to a Patient Assist Call system. See Appendix C -<br />

Installation and Checkout for details.<br />

3) If required, connect a low pressure oxygen source to the ventilator. See Appendix C -<br />

Installation and Checkout for connection and setup details.<br />

4) Connect the Patient Circuit. To keep moisture out of the sense line(s) attached at the patient<br />

wye, be sure to connect the exhalation valve and circuit to the wye so the proximal sense line(s)<br />

are oriented up (see below, flow sensing patient circuit shown). Connect a test lung to the circuit.<br />

5) Set any desired extended features options. For a detailed list of extended features see Chapter<br />

10 - Extended Features.<br />

6) Select the ventilation mode and all controls, including PEEP, to prescribed values. Detailed<br />

procedures for setting each mode are included later in this chapter.<br />

To Turn the Ventilator On<br />

1) Connect the unit to an external source of power. The AC power adapter may be used or the<br />

ventilator may be connected to an external battery.<br />

<br />

<br />

<br />

The External Power LED is lit to indicate the external power source voltage level.<br />

The ventilator begins charging the internal battery from the external source.<br />

The Charge Status LED is lit to indicate the charge progress.<br />

2) Push the On/Standby button and the ventilator will commence operation:<br />

<br />

The On/Standby LED is lit.<br />

21106-001 Rev. E <strong>Operator</strong>’s <strong>Manual</strong> - <strong>LTV</strong> ® <strong>1100</strong> Ventilator 12-3