Pressure Support<br />

This control establishes the target pressure above the PEEP setting for Pressure Support patient<br />

breaths. If Pressure Support is set to dashes “- -“, all patient breaths will be given as Spontaneous<br />

breaths. Inspiratory flow for Pressure Support and Spontaneous breaths is controlled to meet the<br />

patient demand.<br />

NOTE<br />

Pressure Support breaths on the <strong>LTV</strong> ® <strong>1100</strong> are compensated for PEEP. Delivered pressure is<br />

controlled by the Pressure Support setting and is affected by the PEEP setting. For example, a<br />

Pressure Support setting of 20cmH 2 O and a PEEP setting of 10cmH 2 O results in a Peak Inspiratory<br />

Pressure (PIP) of 30cmH 2 O.<br />

REMARQUE<br />

La pression de support des respirations sur le <strong>LTV</strong> ® <strong>1100</strong> sont compensés pour la pression<br />

expiratoire positive (PEP). La pression administrée est contrôlée par le réglage de la pression de<br />

support et elle dépend du réglage PEP. Par exemple, un réglage de la pression de support de<br />

20 cmH2O et un réglage PEP de 10 cmH2O donnent une pression inspiratoire maximale (PImax) de<br />

30 cmH2O.<br />

To set Pressure Support:<br />

1) Push the Pres. Support button.<br />

2) Change the setting using the Set Value knob.<br />

Range: “- -”, 1 - 60 cmH 2 O<br />

Pressure Support breaths may be terminated by flow or by time.<br />

Flow Termination: Pressure Support breaths are flow terminated when the flow decreases to a set<br />

percentage of the peak flow 12 delivered for that breath.<br />

Time Termination: Pressure Support breaths are time terminated 13 when the inspiratory time exceeds<br />

two breath periods, or when the inspiratory time exceeds the set Time Termination Limit 14 before the<br />

flow termination criteria is reached. The Pres. Support display will flash briefly after each time<br />

terminated breath.<br />

The Rise Time profile for Pressure Support breaths may be selected under Extended Features 14 .<br />

12 See Chapter 10 - Extended Features to set the Variable Flow Termination for Pressure Support breaths.<br />

13 Only Pressure Support breaths are flow and/or time terminated (Time Termination must be enabled).<br />

14 See Chapter 10 - Extended Features to set the Variable Time Termination or Flow Rise Time profile.<br />

21106-001 Rev. E <strong>Operator</strong>’s <strong>Manual</strong> - <strong>LTV</strong> ® <strong>1100</strong> Ventilator 6-11