Low Minute Volume<br />

The Low Minute Volume alarm sets the minimum expected exhaled Minute Volume. The exhaled<br />

Minute Volume is recalculated after every breath. If the Minute Volume does not meet or exceed the<br />

Low Minute Volume setting:<br />

A LOW MIN VOL alarm is displayed<br />

The audible alarm is sounded<br />

NOTE<br />

The Low Minute Volume alarm setting is only active and adjustable when Flow Sensing Patient<br />

Circuit mode is enabled. The alarm is inactive and the display window is set to dashes “- -“ when<br />

Non-Flow Sensing Patient Circuit mode is enabled (see Chapter 12 – Operating Procedure for<br />

additional information).<br />

REMARQUE<br />

Le réglage de l'alarme de faible débit-volume n'est actif et ajustable que lorsque le mode de circuit<br />

patient de détection d'écoulement est activé. L'alarme est inactive et la fenêtre d'affichage est réglée<br />

à des tirets « - » lorsque le mode de circuit patient de non détection de circulation est activé<br />

(reportez-vous au chapitre 12 - Procédure d'utilisation pour d'autre information).<br />

To set the Low Minute Volume alarm:<br />

1) Push the Low Min. Vol. button.<br />

2) Change the setting using the Set Value knob.<br />

Range: Off, 0.1 - 99 L<br />

WARNING<br />

Low Minute Volume Control Settings - The Low Min. Vol. control should be set to its highest<br />

clinically appropriate value. If there is a clinical need to set the Low Minute Volume alarm to lower<br />

values or off (“- - -“), perform a clinical assessment to determine if an alternative monitor (i.e. a Pulse<br />

Oxymeter with an audible alarm, or a Cardio Respiratory Monitor) should be used.<br />

AVERTISSEMENT<br />

Réglages du contrôle de volume bas par minute - Le contrôle du volume bas par minute doit être<br />

ajusté à la plus haute valeur clinique appropriée. Si l’alarme de volume bas par minute doit être<br />

ajustée à des valeurs inférieures ou mise à l’arrêt (“- - -“) pour satisfaire aux besoins cliniques,<br />

effectuer une évaluation clinique afin de déterminer si l’utilisation d’un autre moniteur (c.-à-d.,<br />

sphygmo-oxymètre muni d’une alarme sonore ou un moniteur cardio-respiratoire) s’avère pertinente.<br />

21106-001 Rev. E <strong>Operator</strong>’s <strong>Manual</strong> - <strong>LTV</strong> ® <strong>1100</strong> Ventilator 6-5