Variable Flow Termination<br />

Use the Variable Flow Termination to select the percentage of peak flow used for cycling Pressure<br />

Support breaths. Pressure Support breaths are cycled from inspiration to exhalation when the flow<br />

reaches the set percentage of the peak flow, or when flow goes below 2 lpm.<br />

Peak<br />

Flow<br />

<br />

V<br />

Inspiratory<br />

Flow<br />

High<br />

Percentage<br />

of Peak Flow<br />

Peak<br />

Flow<br />

<br />

V<br />

Inspiratory<br />

Flow<br />

Low<br />

Percentage of<br />

Peak Flow<br />

To modify the Variable Flow Termination:<br />

1) Push the Select button while FLOW TERM is displayed.<br />

2) % OF PEAK xx is displayed.<br />

3) Turn the Set Value knob until the desired Variable Flow Termination percentage is displayed.<br />

4) Push the Select button.<br />

Range: 10% to 40%<br />

21106-001 Rev. E <strong>Operator</strong>’s <strong>Manual</strong> - <strong>LTV</strong> ® <strong>1100</strong> Ventilator 10-9