Using the Remote Alarm Cable<br />

Use the Remote Alarm Cable (P/N 19103-001) to connect the <strong>LTV</strong> ® ventilator to third party, single or<br />

dual tone remote alarm systems requiring a normally closed input signal terminated with a 51K ohm<br />

series resistor. Devices connected to the Patient Assist port must be IEC 60601-1-1 certified.<br />

See Chapter 10 - Extended Features, Alarm Operations, for instructions on setting the Patient<br />

Assist Port output signal for use with single or dual tone remote alarm systems.<br />

Because the ventilator does not include an internal series resistor in the Patient Assist output, a<br />

special cable has been designed which incorporates the resistor into the cable assembly itself. The<br />

series resistor allows the remote alarm to detect and report both ventilator alarms and a disconnected<br />

remote alarm cable.<br />

Do not apply more than 120 Volts AC (VAC) to a remote alarm when it is connected to the ventilator.<br />

CAUTION<br />

Remote Alarm – Always verify that the remote alarm properly reports the <strong>LTV</strong> ® ventilator alarms<br />

before use.<br />

Remote Alarm – Always follow the remote alarm manufacturer’s usage and maintenance<br />

requirements to guarantee proper function of the device.<br />

ATTENTION<br />

Alarme à distance – Assurez-vous toujours que l’alarme à distance indique de façon adéquate les<br />

alarmes du ventilateur <strong>LTV</strong> ® avant d’utiliser le ventilateur.<br />

Alarme à distance – Suivez toujours les exigences d’utilisation et d’entretien du fabricant de l’alarme<br />

à distance afin d’assurer le fonctionnement adéquat de l’appareil.<br />

To connect the ventilator to the remote alarm:<br />

1) Plug the cable’s modular jack into the Patient Assist port on the side of the ventilator.<br />

2) If the remote alarm has a female BNC plug, connect the cable directly to the remote alarm's input<br />

cable or connector and twist to secure.<br />