XDCR FAULT<br />

When a transducer autozero test fails, the XDCR FAULT alarm is generated. Transducer autozeros<br />

are scheduled at periodic intervals during ventilator operation. This allows the ventilator to adjust the<br />

zero pressure readings as the ventilator warms up and environmental conditions change. If an<br />

autozero test fails, it will be automatically rescheduled to run again on the next breath. The XDCR<br />

FAULT alarm will remain active until a valid autozero can be done. If the XDCR FAULT persists,<br />

remove the ventilator from service and immediately contact a certified <strong>CareFusion</strong> service technician<br />

or <strong>CareFusion</strong>.<br />

When a XDCR FAULT alarm occurs:<br />

The autozero for the transducer is rescheduled to run again on the next breath.<br />

The XDCR FAULT message is flashed in the display window.<br />

The audible alarm is sounded.<br />

To reset the XDCR FAULT alarm:<br />

1) Push the Silence Reset button twice.<br />

WARNING<br />

XDCR FAULT alarm - Continued operation of the <strong>LTV</strong> ® <strong>1100</strong> ventilator with an activated XDCR<br />

FAULT alarm may result in inaccurate flow and volume measurements. Should this condition occur,<br />

disconnect the patient from the ventilator, provide an alternative method of ventilation and<br />

immediately contact a certified <strong>CareFusion</strong> service technician or <strong>CareFusion</strong>.<br />

AVERTISSEMENT<br />

Alarme XDCR FAULT - L'opération continue du ventilateur de la <strong>1100</strong> <strong>LTV</strong> ® avec une alarme XDCR<br />

FAULT activée peut résulter en mesures de débit et de volume erronées. Si cette condition se<br />

présente, le ventilateur doit être retiré du service, et vous devez immédiatement contacter votre<br />

technicien de service certifié de <strong>CareFusion</strong> ou <strong>CareFusion</strong>.<br />

NOTE<br />

Repeated or continuous XDCR FAULT alarms may indicate a problem with the ventilator that could<br />

prevent the ventilator from performing within its specifications. Discontinue use of the ventilator and<br />

immediately contact a certified <strong>CareFusion</strong> service technician.<br />

REMARQUE<br />

Des alarmes XDCR FAULT répétitives ou continuelles peuvent indiquer un problème qui pourrait<br />

empêcher le ventilateur de fonctionner à l'intérieur des limites spécifiées. Retirer le ventilateur du<br />

service, et contactez immédiatement un technicien de service certifié de <strong>CareFusion</strong>.<br />

9-30 <strong>Operator</strong>’s <strong>Manual</strong> - <strong>LTV</strong> ® <strong>1100</strong> Ventilator 21106-001 Rev. E