Transducer Autozero<br />

Use the Transducer Autozero menu to manually schedule transducer autozeros and to view previous<br />

autozero results. Autozeros are automatically scheduled at appropriate intervals during ventilator<br />

operation, so manual scheduling of autozeros is not commonly performed, but may occasionally be<br />

done.<br />

The menu is set up as follows:<br />

XDCR ZERO<br />

AP xxxx P<br />

FDb xxxx P<br />

FDw xxxx P<br />

FDn xxxx P<br />

NOTE<br />

FDb, FDw, and FDn transducers only available when Flow Sensing Patient Circuit mode is enabled<br />

(see Chapter 12 – Operating Procedure for additional information).<br />

REMARQUE<br />

Les transducteurs FDb, FDw, et FDn ne sont disponibles que lorsque le mode circuit patient de<br />

détection d'écoulement est activé (reportez-vous au chapitre 12 – Procédure d'utilisation pour d'autre<br />

information).<br />

Airway Pressure Transducer Autozero<br />

Use this item to view the Airway Pressure Transducer Autozero results and schedule the Airway<br />

Pressure Transducer Autozero to be run.<br />

To view the Airway Pressure Transducer Autozero results:<br />

1) The previous results, AP xxxx P, are displayed. The final P indicates the previous zero results<br />

were within the required tolerance and the previous autozero passed. If a final F is displayed, the<br />

previous zero results were outside the required tolerance and the autozero failed. An asterisk<br />

indicates that an autozero is scheduled for the next breath.<br />

2) Turn the Set Value knob to display the EXIT option.<br />

3) Push the Select button.<br />

To schedule the Airway Pressure Transducer Autozero:<br />

1) The previous results, AP xxxx P, are displayed.<br />

2) Push the Select button. An asterisk appears, the pass / fail indicator is removed from the display<br />

and the test is scheduled for the next breath.<br />

3) After the autozero is run on the next breath, the new autozero value and the pass / fail indicator<br />

are displayed.<br />

If an autozero fails, it will be automatically rescheduled for the next breath.<br />