DEFAULTS SET<br />

The DEFAULTS SET alarm is generated when the <strong>LTV</strong> ® <strong>1100</strong> ventilator is first powered up after the<br />

SET DEFAULTS 30 option has been used to reset all controls and extended features settings to their<br />

factory-set default values 31 .<br />

<br />

Language, Time/Date Format and Com settings are not reset to default values when the SET<br />

DEFAULTS option is used in Extended Features.<br />

When a DEFAULTS SET alarm is generated:<br />

The DEFAULTS SET message is flashed in the display window.<br />

The audible alarm is sounded.<br />

To reset the DEFAULTS SET alarm:<br />

1) Push the Silence Reset button twice.<br />

2) Select and return the control(s) and extended features settings to the desired settings.<br />

NOTE<br />

Be sure to check all Controls, Alarms and Extended Features options and return them to the desired<br />

settings.<br />

REMARQUE<br />

Assurez-vous de procéder à la vérification de toutes les options de contrôles, d'alarmes et de<br />

caractéristiques étendues, et de les retourner aux réglages souhaités.<br />

30 See Chapter 10 - Extended Features, Set Defaults for additional information.<br />

31 See Chapter 9 - Ventilator Alarms, DEFAULTS for factory-set default values.<br />

21106-001 Rev. E <strong>Operator</strong>’s <strong>Manual</strong> - <strong>LTV</strong> ® <strong>1100</strong> Ventilator 9-9