Test Lung Operations<br />

Symptoms Possible Causes What to Do<br />

Delivered pressure<br />

higher than set<br />

pressure on test lung.<br />

Monitored volumes<br />

very high on test lung.<br />

Pressure> 40 cmH 2 O used<br />

on small test lung<br />

(<strong>CareFusion</strong> or Siemens<br />

190.)<br />

Test lung with small<br />

aperture connected directly<br />

to wye.<br />

Very small ET tube<br />

connected directly to wye.<br />

The compliance characteristics of some<br />

small test lungs (<strong>CareFusion</strong> or Siemens<br />

190) cause incorrect readings when high<br />

pressures are used. For these lungs, use<br />

pressures under 40 cmH 2 O or change to a<br />

larger lung.<br />

Some test lungs have a narrow opening or<br />

a restrictor, which may cause jetting and<br />

cause the flow differential to be read<br />

incorrectly. To reduce the jetting effect,<br />

add a short extension between the test<br />

lung and the wye.<br />

A very small ET tube connected directly to<br />

the wye may cause jetting and cause the<br />

flow differential to be read incorrectly. To<br />

reduce the jetting effect, add a short larger<br />

bore extension between the ET tube and<br />

the wye.<br />

21106-001 Rev. E <strong>Operator</strong>’s <strong>Manual</strong> - <strong>LTV</strong> ® <strong>1100</strong> Ventilator 15-25