Control Test (cont)<br />

3) Test each control by pressing each button, one at a time. When pressed, verify that the name of<br />

the button pressed is displayed in the display window. Control names are as shown in the table<br />

below.<br />

Control<br />

Display Select<br />

Breath Rate<br />

Tidal Volume<br />

Inspiratory Time<br />

Pressure Support<br />

Sensitivity<br />

High Pressure Alarm<br />

Low Peak Pressure<br />

Low Minute Volume 42<br />

Silence Reset<br />

On/Standby<br />

Assist/Control & SIMV/CPAP<br />

<strong>Manual</strong> Breath<br />

Control Lock<br />

Set Value knob rotate Left<br />

Set Value knob rotate Right<br />

PEEP<br />

Display<br />

SELECT<br />

BREATH RATE<br />

TIDAL VOLUME<br />

INSP TIME<br />

PRES SUPPORT<br />

SENSITIVITY<br />

HIGH PRES<br />

LOW PRES<br />

LOW MIN VOL<br />

SILENCE<br />

ON / STNDBY<br />

MODE A/C S/C<br />

MANUAL BRTH<br />

CONTROL LOCK<br />

ROTATE LEFT<br />

ROTATE RIGHT<br />

PEEP<br />

4) Test the Set Value knob by turning it clockwise and counterclockwise. Verify that the direction of<br />

rotation is displayed in the display window.<br />

5) To exit the control test, push the Select button again and the next menu item is displayed.<br />

If the Control Test fails, see Chapter 15 - Troubleshooting for more information.<br />

42 Only available when Flow Sensing Patient Circuit mode is enabled (see Chapter 12 – Operating Procedure for<br />

additional information).<br />

21106-001 Rev. E <strong>Operator</strong>’s <strong>Manual</strong> - <strong>LTV</strong> ® <strong>1100</strong> Ventilator 11-7