Power<br />

Feature Range Tolerance / Indicators<br />

Input Voltage<br />

11 to 15 VDC<br />

External Power<br />

AC Adapter<br />

Input: 100 to 250 VAC,<br />

50 to 60 Hz<br />

Output: 13 VDC<br />

2.5%<br />

Full Power Voltage 11.5 V 2% Green LED<br />

Low Power Voltage < 11.0V and 9.5V 2% Amber LED<br />

External Power Off Voltage < 9.5V 2% LED off, switch to<br />

battery<br />

Hysteresis<br />

Ventilator shall not resume<br />

external power operation unless<br />

voltage is 11.5V<br />

2%<br />

Nominal Current<br />

Draw<br />

Nominal Power<br />

Draw<br />

Startup:<br />

Running:<br />

Startup:<br />

Running:<br />

5.5 amps<br />

3-4 amps<br />

66 watts<br />

36 - 48 watts<br />

Leakage Current<br />

Ground Resistance<br />

Dielectric Strength<br />

Total leakage current to Earth ground for the ventilator with only<br />

approved accessories attached, shall not exceed 500 microAmps<br />

during normal operation, per IEC 601-1.<br />

Total leakage current to Earth ground for the ventilator shall not exceed<br />

one milliAmp when any single fault condition is present, per IEC 601-1.<br />

Total impedance between the ground contact at the inlet power<br />

connector and any accessible metal part shall not exceed 0.1 ohm, per<br />

IEC 601-1.<br />

The ventilator shall be able to survive 1500 volts applied from either<br />

phase of the AC power inlet to Earth ground for a period of one minute,<br />

per IEC 601-1.<br />