7-10 <strong>System</strong> Maintenance<br />

To replace the <strong>7900HT</strong> instrument plate adapter: (continued)<br />

Step Action<br />

4 Place the new plate adapter into the instrument tray with the A1 label in the rear-left<br />

corner (see below).<br />

IMPORTANT Make sure to install the correct version of the plate adapter (384- or<br />

96-well) for the plate format you intend to use. The plate adapters are labeled for<br />

the consumable format they support.<br />

5 Replace and tighten the four screws in the order shown below:<br />

3<br />

2<br />

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24<br />

A<br />

B<br />

C<br />

D<br />

E<br />

F<br />

G<br />

H<br />

I<br />

J<br />

K<br />

L<br />

M<br />

N<br />

O<br />

P<br />

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24<br />

A<br />

B<br />

C<br />

D<br />

E<br />

F<br />

G<br />

H<br />

I<br />

J<br />

K<br />

L<br />

M<br />

N<br />

O<br />

P<br />

1<br />

4<br />

Well A1<br />

Label (384- or 96-well)<br />

IMPORTANT The order in which the screws are tightened is important to ensure<br />

proper alignment of the plate to the sample block within the <strong>7900HT</strong> instrument.