Extracting the<br />

Background<br />

7-16 <strong>System</strong> Maintenance<br />

In this procedure you will extract the calibration values from the background plate<br />

document. Once extracted, the SDS software stores the data as part of the calibration<br />

file located in the Calibration subdirectory of the SDS 2.0 directory.<br />

To extract the background component from the run data:<br />

Step Action<br />

1 From the Analysis menu, select Extract Background.<br />

The software attempts to extract the background signal and displays the success of<br />

the extraction in a dialog box.<br />

Dialog Box Then...<br />

the run is successful<br />

2 From the File menu, select Save.<br />

The software saves the plate document.<br />

3 From the File menu, select Close.<br />

The software closes the plate document.<br />

The raw spectra read from<br />

theBackgroundplate<br />

conform to acceptable limits.<br />

Proceed to step 3.<br />

the run is unsuccessful.<br />

The software has stopped<br />

the extraction because one<br />

or more raw spectra exceed<br />

2500 FSU.<br />

Troubleshoot and decontaminate the sample block as explained in “Background Runs” on<br />

page 8-8.