To adjust the plate sensor switch: (continued)<br />
Step Action<br />
4 Remove the plate and listen for the plate-sensor switch to disengage.<br />
Did you hear the<br />
switch disengage? Then…<br />
No a. MovetheswitchDownafewstepsbyturningthe<br />
thumb wheel in the direction indicated on the arm.<br />
b. Replace the plate within the gripper and listen for<br />
the switch to engage:<br />
– If you do not hear the switch engage, then<br />
remove the plate and repeat steps a and b above.<br />
– If you hear the switch engage, remove the plate<br />
and continue to step a below.<br />
Yes a. Move the switch Up by turning the thumb wheel one<br />
step in the direction indicated on the arm.<br />
b. Replace the plate and listen for the switch to<br />
engage:<br />
– If you hear the switch engage, remove the plate<br />
and repeat steps a and b.<br />
– If you do not hear the switch engage, then you<br />
have successfully identified the zero point of the<br />
plate-sensor switch.<br />
Note At the zero point, one step of the thumb wheel<br />
in the Down direction causes the switch to engage.<br />
5 Once the zero point is established, carefully turn the thumb wheel in the Down<br />
direction the number of steps appropriate for your plate format as indicated below:<br />
Plate Format Turn the thumb wheel in the Down direction…<br />
96-Well 20 steps<br />
384-Well 15 steps<br />
Note If you lose count, begin again from step 4 and identify the zero point for the<br />
switch.<br />
6 Test the adjustment as explained on page 7-30.<br />
<strong>System</strong> Maintenance 7-29