To determine the location of the contamination on the sample block: (continued)<br />

Step Action<br />

Example: (continued)<br />

b. The raw data from<br />

the selected wells<br />

does not contain<br />

the peak.<br />

The contaminated<br />

well must be in the<br />

last four wells of<br />

columns 7-12.<br />

c. The raw data from<br />

the selected wells<br />

contains the peak.<br />

The contaminated<br />

well must be in<br />

the last four wells<br />

of columns 10-12.<br />

d. The raw data from<br />

the selected wells<br />

contains the peak.<br />

The contaminated<br />

well must be in<br />

the last two wells<br />

of columns 10-12.<br />

e. By selecting each<br />

of the wells from<br />

the last two wells<br />

of columns 10-12,<br />

thelocationofthe<br />

contaminated<br />

well (G10) is<br />

determined.<br />

6 Repeat step 4 until you identify the location of each contaminated well.<br />

7 Decontaminate the sample block as explained in “Decontaminating the Sample<br />

Block” on page 7-11.<br />

8 Run a background plate to confirm that the contaminants have been removed.<br />

If the contamination is present after running the background plate for a second time,<br />

the background plate is likely to be the source of contamination.<br />