Cleaning and Replacing Gripper Finger Pads<br />

When to Perform The adhesive used to affix bar code labels to certain brands of microplates can build<br />

up on the gripper pads of the Zymark Twister Microplate Handler. Over time, the<br />

residue can cause the gripper pads to stick to the microplates while handling them,<br />

causing misfeeds. To prevent buildup, inspect the gripper pads monthly and clean or<br />

replace the pads as needed.<br />

Materials Required The following materials are required to replace the finger pads:<br />

Cleaning the<br />

Finger Pads<br />

Replacing the<br />

Finger Pads<br />

Material Part Number<br />

Finger Pad Replacement Kit, containing 10 finger pads 4315472<br />

Flat-blade screwdriver, small —<br />

Phillips head screwdriver, small —<br />

Isopropanol in a squeeze bottle —<br />

! WARNING CHEMICAL HAZARD. Isopropanol is a flammable liquid and vapor. It may<br />

cause eye, skin, and upper respiratory tract irritation. Prolonged or repeated contact may dry<br />

skin and cause irritation. It may cause central nervous system effects such as drowsiness,<br />

dizziness, and headache, etc. Please read the MSDS, and follow the handling instructions.<br />

Wear appropriate protective eyewear, clothing, and gloves.<br />

To clean the finger pads, wipe each pad thoroughly with Isopropanol until the residue<br />

has been resolved. If the pads appear rough or the adhesive cannot be removed,<br />

replace the pads as described below.<br />

To replace the finger pad(s):<br />

Step Action<br />

1 Using the Phillips-head screwdriver, remove the two small Phillips-head screws<br />

fromthefingersoneachsideofthegripper,thenremovethefingers.<br />

Note Move the plate handler arm into any position where it is easy to access the<br />

screws.<br />

2 Using a small flat-blade screwdriver, pry the worn finger pads off the fingers.<br />

Note The manufacturer recommends replacing all finger pads at the same time.<br />

3 Clean any residual adhesive off the fingers using isopropanol.<br />

4 Remove a replacement finger pad from the paper backing, and place the finger pad<br />

on the appropriate finger position.<br />

5 Repeat for the remaining finger pads.<br />

6 Install the fingers with the fingers pointing down and the finger pads toward the<br />

center of the gripper.<br />

7 Insert the screws into the fingers and tighten.<br />

Note The screws do not automatically align the grippers. Make sure that the<br />

finger pads are making good contact with the plate when the arm grips a plate.<br />

<strong>System</strong> Maintenance 7-43