7-14 <strong>System</strong> Maintenance<br />

To prepare a plate document for the background run:<br />

Step Action<br />

1 Launch the SDS software.<br />

2 From the File menu, select New.<br />

The New Document dialog appears.<br />

3 Configure the New Document dialog box as follows:<br />

Drop-Down List Select<br />

Assay Background<br />

Container <br />

Template Blank Template<br />

4 If the background plate is labelled with a bar code, click the Barcode text field and<br />

scan the bar code number using the hand-held bar code reader.<br />

5 Click OK.<br />

The software creates a plate document with the attributes for a background run.<br />

Note Do not modify the background plate document. The method for a<br />

Background run is hard-coded into the SDS software and consists of a single hold<br />

at 60 °C for 2 min. Because the plate contains only deionized water, the plate<br />

document does not require sample or detector labels.<br />

6 Save the plate document as follows:<br />

a. From the File menu, select Save.<br />

The Save dialog appears.<br />

b. Click the File name text field, and type Background_.<br />

For example, the file name for a background plate run on May 31, 2001 would<br />

be: Background_053101<br />

c. Click Save.<br />

The software saves the plate document. The software is now configured for the<br />

Background run.<br />

7 Prepare and run the plate as explained on page 7-15.