About the Setup<br />

Table File Format<br />

Setup Table Elements<br />

This section explains the elements of setup table files shown on the previous page.<br />

The following table describes the conventions used in the rest of this section.<br />

Format/Symbol Definition<br />

courier Text appearing in bolded courier font must be applied to a setup<br />

table file exactly as appears in this document.<br />

italic Text appearing in italic courier font must be substituted with<br />

custom values when applied to a setup table file.<br />

[ required text ] Text appearing between brackets is required information in setup<br />

table files. All information within the brackets must be present in<br />

the setup table file for the SDS software to import it.<br />

{ required text } Text appearing between braces is optional in setup table files.<br />

The tab character (the equivalent of pressing the Tab key)<br />

The carriage-return character (the equivalent of pressing the<br />

Enter key)<br />

IMPORTANT To guarantee a successful importation of the setup table file into a plate<br />

document, the file can must contain all of the sections in the following table in the order that they<br />

appear in this document.<br />

Number Contents Description<br />

1 File Version<br />

Format:<br />

This line defines the version of SDS Assay Plate File format used to generate the document.<br />

[ *** SDS Setup File Version version number ]<br />

Example:<br />

*** SDS Setup File Version 2<br />

2 Plate Size<br />

Format:<br />

This line defines the number of wells in the plate modeled by the file (384 or 96).<br />

[ *** Output Plate Size number of wells ]<br />

Example:<br />

*** Output Plate Size 384<br />

3 Plate ID<br />

Format:<br />

This line defines the ID of the Assay Plate. Normally this will be a bar code that is printed on<br />

the plate.<br />

[ *** Output Plate ID plate id ]<br />

Example:<br />

*** Output Plate ID 384N75822034<br />