To define the positions of the remaining stacks: (continued)<br />

Step Action<br />

8 Using the Vertical Positioning commands, raise and lower plate handler arm several<br />

times to check the alignment.<br />

9 Click Rotary Offset, andclickYes.<br />

The software records the rotary position for the Zymark position 5 (input stack 2).<br />

10 Repeat steps 1–8 for input stacks 3 and 4 to define Rotary Offset values for the<br />

remaining positions 6 and 7:<br />

Zymark position 7<br />

(input stack 4)<br />

Zymark position 6<br />

(input stack 3)<br />

11 Exit the Zymark Twister Software.<br />

a. Click Main Menu.<br />

The software displays the main menu.<br />

b. Click Exit.<br />

The software closes.<br />

12 Note A bug in the Zymark Twister Software can cause portions of the program to<br />

persist in memory even after the software has been closed. Because the Zymark<br />

Twister Software conflicts with the SDS software, these residual elements must be<br />

closed within the Windows Task Manager before continuing.<br />

Confirm that the Zymark Twister Software has closed by viewing the Task Manager.<br />

a. Press the Ctrl + Alt + Del keys in unison.<br />

The Windows Security dialog box appears.<br />

b. Click Task Manager.<br />

The Task Manager dialog box appears.<br />

c. Confirm that the Twister software has closed by looking for the Twister software<br />

entry in the Task list. If the software is still running, click the software entry and<br />

click End Task to exit the software.<br />

d. From the File menu, select Exit to quit the Task Manager.<br />

13 Replace the covers for the fixed-position bar code reader and the underlying<br />

platform (removed in step 1 on page 7-32).<br />

GR2009<br />

Fixed-position bar code reader<br />

and underlying platform covers<br />

<strong>System</strong> Maintenance 7-39