23.07.2013 Views

ABI Prism® 7900HT Sequence Detection System ... - OpenWetWare

ABI Prism® 7900HT Sequence Detection System ... - OpenWetWare

ABI Prism® 7900HT Sequence Detection System ... - OpenWetWare

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

Writing on the<br />

Reaction Plates<br />

Fluorescent<br />

Contamination on<br />

the Plates<br />

8-6 Troubleshooting<br />

Drops Drops of reagents that cling to the sides of the wells may not contact the thermal<br />

cycler sample block and consequently may not amplify. If the drop slides into the mix<br />

during PCR, then the amplified products will become diluted and the final result will be<br />

less than replicate wells that did not have drops. Therefore, carefully monitor the<br />

reaction plate as it is being transferred into the thermal cycler or <strong>7900HT</strong> instrument. If<br />

you observe any drops, take steps to remove them, such as centrifugation.<br />

Do not write on any surface of the Optical 384/96-Well Reaction Plates or the Optical<br />

Adhesive Covers. The fluorescent properties of the ink can potentially affect the<br />

fluorescence emission from the plate and alter the results. Instead, note the contents<br />

of each well on a sheet of paper, or on a printout of the sample setup.<br />

Many compounds found in laboratories are fluorescent. If they come in contact with<br />

certain optical surfaces, such as the optical adhesive covers, the fluorescent results<br />

may be affected. For example, it has been noted that the powder used to lubricate the<br />

insides of plastic gloves often contains fluorescent compounds. Use only powder-free<br />

gloves and do not needlessly touch the reaction plates or optical adhesive seals.<br />

Errors Human errors from time to time are inevitable, such as pipetting into the wrong well, or<br />

making a dilution mistake.<br />

Human error can be reduced in the following ways:<br />

♦ Perform the assay in a systematic fashion. For example, the pattern of sample<br />

positions should be simple (such as avoid putting gaps in the rows).<br />

♦ When pipetting the master mix, look directly down into the reaction plate so that<br />

you can verify the transfer of the solution.<br />

♦ If adding a small-volume reagent, such as template, place the drop of liquid on the<br />

side of the well. Briefly tap or centrifuge the plate afterwards to bring the droplet<br />

down into the well.<br />

♦ After all pipetting is complete, visually inspect all the wells to confirm the presence<br />

of the reagent drops. Tapping or centrifuging the reaction plate will cause all the<br />

drops to slide down into the wells simultaneously.<br />

♦ When making serial dilutions, be sure to change the pipet tip after each dilution<br />

step.<br />

♦ Visually inspect the liquid volumes being pipetted to verify that the volume is<br />

approximately correct. A common mistake is using the wrong pipettor volume<br />

setting (such as setting 20 µL instead of 2.0 µL).<br />

♦ Visually inspect the volumes of the completed reactions, looking for any wells that<br />

have volumes that do not match those of the other wells.

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!